Instructions for use DIAGNOcam Vision Full HD
7 Reprocessing steps in accordance with ISO 17664 | 7.1 Overview of the reprocessing options
7 Reprocessing steps in accordance with ISO 17664
7.1 Overview of the reprocessing options
Components Single use
or Multi-
person use
Manual
cleaning
Manual dis-
infection
Automated
cleaning
Automated
disinfection
Sterilisation
DIAGNOcam
Vision Full HD
Multi-Person
use
Yes
Yes
No
No
No
Tip Vision Full
HD
Multi-Person
use
Non-fixing
pre-cleaning
No
Yes
Yes
Yes
Fluorescence
Cover Vision
Full HD
Multi-Person
use
Non-fixing
pre-cleaning
No
Yes
Yes
Yes
Cradle Vision
Full HD
Multi-Person
use
Yes
Yes
No
No
No
Wall Mount
Vision Full HD
Multi-Person
use
Yes
Yes
No
No
No
Foot Control
Vision Full HD
Multi-Person
use
Yes
Yes
No
No
No
Sheath Vision
Full HD
Single Use
No
No
No
No
No
7.2 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
▶
Pull off the USB plug of the medical device.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶
Remove gross soiling immediately.
▶
Do not place in solutions or similar substances.
7.3 Disassembly
WARNING
Incomplete reprocessing.
Infection hazard.
▶
To ensure complete reprocessing of all parts, camera, tip and Fluorescence
Cover need to be removed before reprocessing.
37 / 52