7-2 DIAGNOSTICS
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
Help Code
or Error
Message
Description
Possible Causes
And
Corrective Action
current
fl
owing in the
ground connection of the
serial interface between
the unit and the Computer.
Shows up only in remote
RS-232 operation.
Display says HLP 8 for 5
seconds.
recovers from error condition if
problem goes away.
Make sure all of the equipment
has adequate grounding.
HLP 10
Start failed.
The
discharge did not start.
Press and release the STANDBY/
ENABLE button or remove the
remote enable signal to clear the
code. The unit remains in
STANDBY until STANDBY/EN-
ABLE is pressed again or the
remote enable signal is reapplied.
Anode has an insulating coating
and needs to be cleaned.
Hollow Cathode did not start. In
this case, HLP 10 is also dis-
played on the Keeper Power Sup-
ply. Review the HLP 10 message
for the keeper supply below in
table 7.2.
For initial operation at a new
operating condition, the gas
fl
ow
may be too high or too low. De-
crease or increase
fl
ow.
Open or shorted discharge (an-
ode) cable. Inspect in vacuum
discharge cable and replace if
necessary. Not all damage to the
discharge cable may be appar-
ent; in this case replacing the
cable may be the only evidence
the cable was faulty.