6-22 REMOTE/AUTOMATED CONTROL
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
Keeper Amps = 1.500
Bias Volts = 38.000
Bias Amps = 3.000”
In ‘Terse’ mode the response will be a comma delimited
string of values only in the same as order shown above for
the ‘Verbose’ mode.
6.4.4.11 CFG? – Query current system con
fi
guration
The CFG? Command is a query of what the Auto-Controller thinks the
current con
fi
guration is based on the initial setup.
Command:
CFG?
In VERBOSE, the responses are:
“Filament”
“Hollow Cathode with BV”
“Hollow Cathode without BV”
“Gas
Only”
“Unknown”
In ‘TERSE’, the responses are: “0”, “1”, “2”, “3” or
“4”,
respectively.
6.4.4.12 LRN – Learn Command or Learn Query
The current status of LRN can be queried any time. The COM must be
active to set the LRN
fl
ag. Setting the LRN
fl
ag with COM in-active is
ignored. Setting the LRN mode using the serial command temporarily
overrides the LRN mode setting established in ‘setup’. The value in
‘setup’ is restored on a power down/up or reset.
When LRN is set to on, each program will remember the last running gas
fl
ow when the program is disabled and use that gas
fl
ow the next time
the program is enabled. When LRN is turned off, the program will always
start with the originally programmed gas
fl
ow.
Command: LRN:<parameter>
Command: LRN?
Parameter/Response: 1 = ON, 0 = OFF
<parameter> can be separated from the command by l or less spaces.