77
Error / warning
codes listing
Error
Code
Description
D1
Door motor slipped.
D2
Over-temperature door motor drive.
D3
Pre-Warning – Over-temperature door motor drive.
D4
Short to ground - Loop A Door motor.
D5
Short to ground - Loop B Door motor.
H1
Element #1 low power detected.
H2
Element #2 low power detected.
H3
Element #3 low power detected.
H4
Element #4 low power detected.
H5
Element #5 low power detected.
H6
Element #6 low power detected.
H7
Element #7 low power detected.
H8
Broken SSR or leak current to Ground.
H9
Two or more heating elements show low power.
H10
Max Power exceeded. Short-circuit or broken element.
H11
Furnace: overheating.
H12
Furnace: temperature is not increasing.
H13
HMI detected: overheating.
H14
Low voltage detected.
M1
Motor controller: communication problem.
P1
Pause.
P2
Preheating.
R1
Rocking motor slipped.
R2
Over-temperature rocking motor drive.
R3
Pre-Warning - Over-temperature rocking motor drive.
R4
Short to ground - Loop A Rocking motor.
R5
Short to ground - Loop B Rocking motor.
S1
Sliding motor slipped.
S2
Over-temperature sliding motor drive.
S3
Pre-Warning – Over-temperature sliding motor drive.
S4
Short to ground - Loop A Sliding motor.
S5
Short to ground - Loop B Sliding motor.
T1
Thermocouple not connected.
T2
Thermocouple error.
T3
Temperature controller: communication error.
T4
Temperature controller: error.
T5
Temperature controller: not read.
T6
Temperature controller: timeout.