57
fasten the collar onto the rocking shafts, and so must not be un-tightened at
this point.
4.
One must start by removing the Hoop:
4.1
Locate and un-tighten the screw that is all by itself (not in a group of 4).
You do not need to fully remove the screw.
4.2
You can now gently pull on the hoop to remove it. Notice that it features
a notch where the holding screw was located.
5.
Now, let us remove the fork:
5.1
Locate and un-tighten the two collar screws that are closer to the front
of the unit. You do not need to remove the screws.
5.2
You can now gently pull on the fork to remove it.
6.
Replace or clean any damaged or flux-soiled parts.
Crucible
holder
installation
In this section, you will find essential information needed to properly re-install the
crucible holder. Please follow the instructions carefully, or pouring issues could
occur.
1.
Manually rotate the crucible rocking collar, so that its four (4) grouped screw
head point upwards.
2.
Carefully look at the fork, and you will notice that it is not symmetrical. One
prong of the fork has a longer transversal branch. This longer branch goes to
the right.
3.
Now, fully insert the main stem of the fork in the horizontal slot; remember to
keep the longer branch on the right.
4.
Tighten the two screws that will compress and hold the fork stem. Make sure
to tighten both approximately at the same rate, so as to close the gap
uniformly.
5.
Now, rotate the collars about 90° counterclockwise. A lone screw head (for the
hoop) should now be pointing upwards.
6.
Insert the hoop in its slot, making sure to align its notch with the screw, i.e.
towards the right.
7.
Gently tighten the hoop screw. Note that the screw should not be compressing
anything; it is the threaded body of the screw in the hoop’s notch that will
prevent the latter from coming off.
Mold holder
removal
Removing the mold holder(s) is required when preparing solutions.