Electromagnetic compatibility
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR
46
9.3 Table 1 – Compliance level for immunity tests
Guidelines and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user of the
product should make sure that it is used in such an environment.
Immunity tests
EN/IEC 60601 test level Compliance level
Electromagnetic envi-
ronment – guidelines
Electrostatic dis-
charge (ESD) acc.
to IEC 61000-4-2
± 8 kV contact discharge
± 15 kV air discharge
± 8 kV contact discharge
± 15 kV air discharge
Floors should be made of
wood, concrete, or cov-
ered with ceramic tiles. If
floors are covered with
synthetic material, the rel-
ative humidity must be at
least 30%.
Electrical fast tran-
sients/bursts acc.
to IEC 61000-4-4
± 2 kV for power lines
± 1 kV for input and out-
put lines
100 kHz repetition
± 2 kV for power lines
± 1 kV for input and out-
put lines
100 kHz repetition
The power supply quality
should be that of a typical
commercial or hospital
environment.
Surges acc. to IEC
61000-4-5
±1 kV voltage outer con-
ductor – outer conductor
±2 kV voltage outer con-
ductor – ground
±1 kV voltage outer con-
ductor – outer conductor
±2 kV voltage outer con-
ductor – ground
The power supply quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations
acc. to
IEC 61000-4-11
Voltage dip:
Dip to 0% for 1 cycle at
0° phase angle
Dip to 70% for 25/30 cy-
cles at 0° phase angle
Dropout to 0% for 0.5
cycles @ 0°, 45°, 90°,
135°, 180°, 225°, 270°,
and 315° phase angles
Voltage interruption:
100% for 250/300 cy-
cles
Voltage dip:
Dip to 0% for 1 cycle at
0° phase angle
Dip to 70% for 25/30 cy-
cles at 0° phase angle
Dropout to 0% for 0.5
cycles @ 0°, 45°, 90°,
135°, 180°, 225°, 270°,
and 315° phase angles
Voltage interruption:
100% for 250/300 cy-
cles
The power supply quality
should be that of a typical
commercial or hospital
environment. If the user
of the product requires
continued operation in
the event of interruptions
to the power supply net-
work, it is recommended
that the product be oper-
ated with an uninterrupt-
ible power supply or a
battery.
Magnetic field at
the power fre-
quency (50/60 Hz)
acc. to IEC
61000-4-8
30 A/m at 50 Hz / 60 Hz
30 A/m at 50 Hz / 60 Hz
If image distortion occurs,
it may be necessary to in-
stall the product further
from sources of electro-
magnetic fields or to in-
stall magnetic shielding.
Before the product is in-
stalled, the electromag-
netic field should be mea-
sured to ensure that it is
sufficiently low.
Immunity test acc.
to IEC 61000-4–3
for radiated, radio-
frequency electro-
magnetic fields
3 V/m 80 MHz to
2.7 GHz
3 V/m 80 MHz to
2.7 GHz