Preparation
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR
24
5 Preparation
5.1 Unpacking the product
1.
Carefully remove the product and accessories from the packaging.
2.
Check the delivery for missing items and possible damage.
3.
In the case of damage, hidden defects, and short deliveries, document their nature and extent
and contact the manufacturer or supplier immediately.
4.
Keep packaging for further transport.
5.2 Setting up the product
WARNING
Blocked air inlets and air outlets! Risk of fire!
If the air inlets and outlets are blocked, the product may overheat. This can cause the product to fail
and start a fire. Users, patients, and third parties may be injured.
Remove any blockages in the air inlets and outlets when setting up the product.
CAUTION
Breakable glass! Risk of injury!
The front glass will break if the product is dropped or sustains a significant impact. Patients, users, or
third parties can injure themselves on broken glass.
Do not touch broken glass.
Do not continue to use the product.
Do not touch the glass parts of the product.
Remove small glass parts from the product.
Have the product repaired by qualified service personnel.
The product can be operated free-standing, on a video cart, or in a rack.
1.
Place the product on a horizontal, flat surface.
2.
Keep the product out of reach of patients.
3.
Keep air inlets and outlets clear.
4.
Ensure adequate air circulation.