Application
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR
37
–
30 days
–
90 days
6.10.2 Labeling
1.
Specify whether the digital recordings or prints are to be labeled with patient information or not.
2.
Select one of the following options:
–
No patient data in the patient folder (selected by default)
–
Patient data in the patient folder
–
Patient data on prints
6.10.3 Export mode
Export mode specifies the permitted form of data export. The data can be saved on the USB storage
medium or transferred to a network storage path. The two storage options can also be permitted in
parallel.
Select Export mode:
–
No export
–
USB (or USB anonymous) export (selected by default)
–
Network export
–
Network and USB export
6.11 Network communication
The product can be operated in network mode in order to place recorded data at a user-defined
storage location.
The data can be stored on an FTP server or on a worklist server.
The data generated are temporarily stored in the internal memory with encrypted assignment and
transferred to the defined server. If the network is not available, all data remain stored internally. The
data are automatically sent to the selected server as soon as the network becomes available.
6.11.1 FileShare mode
In FileShare mode, a new patient data set is created in the patient area or selected from an existing
list. The video data and images generated during the treatment are automatically transferred to the
defined server directory.
A sub-directory is created for each patient. As soon as a file has been successfully transferred from
the internal memory to the server directory and the treatment is complete, the local copy is deleted. If
the network connection is interrupted during treatment, all data are automatically transferred as soon
as the network connection becomes available again.