16 KANE 458s MANUAL
PRINTER TYPE
IR PRINT
Select, KMIRP, IRP-2
02 REF
02 REF
Use for “Normalised” readings -
Default
set to 3%, can adjust up or down
LANGUAGE
LANG
Selects your required language
CODE
CODE
Password protected for
authorised service agents only -
Default to 000000
KANE LINK
You can wirelessly connect optional KANE LINK devices to your
analyser. Once connected, they stay connected until you use KANE
LINK to remove them.
If on, they replace or add to you analyser measurements your analyser
makes.
See page 29 to add or remove optional KANE LINK devices
MEASURING FLUE GASSES
After countdown is finished and your analyser is ready to use, put its
flue probe into the appliance’s sampling point. The probe tip should be
at the centre of the flue. Use the flue probe’s depth stop cone to set in
position.
With balanced flues, make sure the probe is positioned far enough into
the flue so no air can “back flush” into the probe.
WARNING
Ensure the flue probe handle does not get hot!