26 KANE 458s MANUAL
ROOM CO TESTING
Rotate dial to Room CO to measure & record room CO readings for up to
30 minutes.
Use to select test type from the following options:
TEST TYPES
TEST TYPE
DURATION
LIMITS/ALARM
LEVELS
GENERAL
15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
SWEEP TEST
2 minute test with max read-
ing stored at end
LIMIT = 10ppm
ALARM = 30 ppm
MIGRATION TEST 15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE C SEALED
APPLIANCE
15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE B BOILER
OPEN FLUE
15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE A COOKER
30 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE A WATER
HEATER
5 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE A SPACE
HEATER
30 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm