ULTRAFLOW® 54/34
Kamstrup A/S · Technical description · 5512-2464_A1_GB_08.2019
71
9.2
Electrical connection
Connection via 3-wire cable from ULTRAFLOW®
Yellow
Signal
Red
Supply
Blue
Ground
Supply
3.6 VDC ±0.1 VDC
Output using Pulse Transmitter/Pulse Divider with galvanically separated output module (Y=2)
Type
Open collector. Two-wire or three-wire connection is possible via the built-
in 56.2 kΩ pull
-up
resistor.
Module Y=2
OC and OD
(OB) Kam
Max input voltage
6 V
30 V
Max input current
0.1 mA
12 mA
ON condition
U ≤ 0.3 V @ 0.1 mA
U
CE
≤ 2.5 V @ 12 mA
OFF condition
R ≥ 6 MΩ
R ≥ 6 MΩ
Table 29
Figure 60. Block diagram for galvanically separated output module (Y=2).