MULTICAL® 801
104
Kamstrup A/S · Technical Description · 5512-571_R1_GB_11_2016
13.2.2
Technical data
Power supply (97-98):
5…30 VDC
Current consumption:
Max. 5 mA
Volume simulation:
Max. 128 Hz for CCC=1xx and 4xx (ULTRAFLOW
)
Max. 1 Hz for CCC=0xx (Reed contact)
HF-energy output (13-12):
Open collector, 5…30 VDC max. 15 mA
Pulse frequency (13-12):
Max. 32 kHz as burst per integration
Data interval:
About 7 s.
Time-out in case of missing data:
About 35 s.
13.3
True energy calculation
During test and verification the heat meter’s energy calculation is compared to the ”true energy” calculated
according to the formula of EN 1434-1:2004 or OIML R75:2002.
An energy calculator like the one shown below can be ordered from Kamstrup A/S:
The true energy at the most frequently used verification points is indicated in the table below.
T1
[°
C
]
T2
[°
C
]
∆Θ
[
K
]
Inlet
[
Wh/0.1 m
3
]
Outlet
[
Wh/0.1 m
3
]
42
40
2
230,11
230,29
43
40
3
345,02
345,43
53
50
3
343,62
344,11
50
40
10
1146,70
1151,55
70
50
20
2272,03
2295,86
80
60
20
2261,08
2287,57
160
40
120
12793,12
13988,44
160
20
140
14900,00
16390,83
175
20
155
16270,32
18204,78