EIASMANUAL 0918
KAM CONTROLS, INC.
4
Before installing the E-IAS™ Sampler, thoroughly inspect to ensure that unit was not damaged during transit.
KAM CONTROLS recommends that the E-IAS™ Sampler be installed 2 to 4 pipe diameters downstream of a
KAM
®
static mixer and in conjunction with a KAM
®
ML™ Measurement Loop. At the time of installation the
pipeline pressure needs to be reduced to under 100 psi in order to safely install the E-IAS™ Sampler.
The KAM
®
E-IAS is shipped with the motor and housing separate from the probe body. KAM recommends full
installation of probe (including insertion on insertable models) prior to attaching motor housing to probe body. If
motor housing is attached at time of installation, detach prior to installation by removing the (4) hex screws at the
bottom of the housing with a 3/16 Allen wrench. (Fig. 3-1)
PRIOR TO INSTALLATION
FIXED INSERTION MODELS INSTALLATION
I N S TA L L AT I O N
FIG. 3-1
Prior to installation,
remove hex screws with a 3/16 Allen wrench
In horizontal lines, the E-IAS™ Sampler should be installed with the probe horizontally oriented at 3 or 9 o’clock.
KAM recommends using liquid pipe sealant rather than Teflon tape on all threaded connections.
Insertion distance/shaft length is set at factory. After installation of probe body on the line, attach the motor housing
to probe body using four hex screws and a 3/16 Allen wrench. (Fig. 3-1).
Follow wiring instructions on page 10 of this manual.