Chapter Five System Installation and Check
5.1 System installation
Please carefully read through and fully understand the use-method before installing the system, and
check the goods for its completeness according to the packing list furnished. This system provides the
following several usages, for the user to select:
1. Place the instrument on a desktop to use
Warning:
1.
If breakers and fuse of the mains power socket are identical to those of this system, and they
are used to control the current for equipment like life support system, the system shall not
be connected to such power supply socket as it may cause breaker or fuse to trip and cut off
the power supply to the entire premise in case of malfunction or over current or transient
current generated at power-on with this ultrasonic system.
2.
All plugs of instruments of this system shall be connected into the power socket with
protectively earth on the wall and the socket must meet the requirement of power rating of
instrument. Multiple portable socket-outlets can not be used for the system.
3.
Equipment that connects the signal input part or signal output part must only connect the
accessories authorized in this manual, and connect the equipment that complies with the
respective IEC standards (IEC 60950 for data processing equipment and IEC 60601-1 for
medical equipment). If in doubt, consult the technical service department or your local
representative.
4.
When this system is installed or used in the patient environment, try to avoid the patient
touching the system. If system is with some unknown defects, it may cause danger of electric
shock.
5.
If the integrity of the external protective conductor in the installation or its arrangement is
in doubt, equipment shall be operated from its internal electrical power source.
Warning:
1.
When instrument works abnormally, do stop working, turn off the power and check the
reason, then contacts the KAIXIN Company about it.
2.
Turn off power and pull out of the plug from socket after each ultrasonic diagnostic
operation.
3.
It is forbidden to drag and press the power and probe cables emphatically; regularly inspect
whether there is pull-apart and bareness, if there is the phenomena like this, turn off power
supply immediately, stop using it and change it for new one.
4.
It is forbidden to load and unload the probe or move the instrument in galvanic to avoid
danger of safety.
5.
Pull out of the plug from socket after operation in thunderstorm weather to avoid the
instrument being damaged by lightening.
6.
If the temperature changes greatly in short time will cause vapor recovery inside of
instrument, the case may damage the instrument.
7.
The instrument is turned off completely only by disconnecting the power supply from the
wall socket.
Warning:
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.