Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet) User’s Manual V1.03
- 7 -
Upper cover of battery storage
Battery
Push button of battery storage
Fig. Install, remove the battery
5.4 Shutter release installation
Insert shutter release into shutter release interface (Freeze interface) as shown in the figure.
Shutter release
In the operation, you can
"freeze" or "unfreeze" images
using shutter release.
5.5 Video recorder installation
1. Turn off the system, connect the equipotential terminal ( ) of the video recorder to the earthing;
2. Connect one end of the video connection line to the video recorder and the other end to the video
output interface on the left of main unit;
3. Insert one end of power plug (jack) of the video recorder to the power input socket of the video
recorder, the other end to the power supply socket.
5.6 Connect to power
1. Connect to the power adapter
Insert the output plug of adapter into power input interface of main unit.
2. Connect to the main power supply
Insert the power plug (jack) furnished with the machine into power input socket of the power adapter,
the other end to the main power supply socket. The instrument uses three-core power supply. It
connects with the protective earth line when power plug inserts into its socket.
Warning:
1. Adapter has no switch. The isolation of the system with the MAINS used to unplug the
adapter as the intended isolation means.
2. The device should be used only with power adapter provided by Kaixin Company.
3. To avoid damaging power adapter or harming people by unexpected fallen, make sure the
power adapter is placed on the leveled desk.
4. The operator must not touch signal input/ signal output and patient simultaneously.