7 User maintenance
7.1 Cleaning
The G85 is a low maintenance unit designed to be both easy to use and clean, plus it is
durable due to its construction.
All surfaces should be cleaned with an 70% alcohol solution on a clean cloth or lint-free
paper towel.
•
Clean all surface components by wiping gently with gauze.
•
To remove fingerprints or oil smudges, wipe with gauze slightly moistened with a
mixture of ether (70%) and alcohol (30%).
Note
Since solvents such as ether and alcohol are highly flammable, they must be
handled carefully. Be sure to keep these chemicals away from open flames or
potential sources of ignition – electrical equipment must be switched off.
Also remember to always use these chemicals only in a well ventilated room.
•
If smudges are difficult to clean, wipe them with a soft cloth slightly moistened with
a diluted neutral detergent.
•
The heated area is able to withstand some water spills. However, precautions are
still necessary if water is spilled on the surface. Unplug the power cord, and then
wipe dry with a dry cloth immediately.
Note
If water gets inside the unit, contact your K-SYSTEMS Kivex Biotec
representative to check electrical safety.
•
Humidification bottle can be put in an autoclave.
7.2 Calibration
For optimal performance the unit should be calibrated once a year by authorized K-
SYSTEMS Service provider.