5.8 Temperature/Flow alarm
The G85 has one alarm which is activated on two events:
•
Temperature deviations (high or low)
•
Gas flow
5.8.1 Temperature alarm
The temperature alarm indicates too high or too low temperature.
The alarm will be activated if the temperature rises or falls more than 0.5° C from the
setpoint.
There will be an acoustic signal and the red LED light in the alarm key will be activated.
Action
Key
Press setpoint key and alarm key to mute
the acoustic signal.
Press setpoint key and alarm key once
again to bring the signal back on.
Note
The red LED and the acoustic signal will turn off if the temperature or gas flow
stabilizes again to the setpoint level.
5.8.2 Gas flow alarm
The gas flow alarm will get activated, if the flow is higher than 29 l/h.
The display will then show
“hi”
.
The alarm will also be activated if the flow is “on”, and the heat “off”.
In this case the display will show
“f.on”
.
Note The maximum flow is 29 l/h. The G85 will not be damaged by a higher flow as
long as the gas flow is kept below 100 l/h.