Instruction Manual - G85
28
8 Trouble shooting
Symptom
Cause
Action
Temperature alarm
is on.
Temperature more
than ±0.5° C from
set temperature.
Check the set
temperature. Wait
for the temperature
to stabilize.
No heating.
Setpoint below
ambient.
Change setpoint
(eg. 37° C).
No mains supply.
Plug in the power
connector.
Power connector,
not connected.
Reconnect power
connector.
Display remains off
when “HEAT” key is
activated.
Defective “HEAT”
key.
Replace the
keyboard.
Flow alarm is on
slow acoustic signal
sounds and the
display shows “-hi-“.
The flow is higher
then 29 l/h.
Reduce the flow.
Flow alarm is on.
Fast acoustic signal
sounds and display
shows “f.on“.
The flow is on and
the heat is off
Turn the flow off.