H2 Stationary GrainGage User’s Guide
54
H2 Stationary GrainGage Calibration
If desired, first tap the gear icon on the right for more information about this moisture curve.
Note: Only adjust these settings if a HarvestMaster Field Service Engineer tells you to.
4.2.4 Test and Tune a Moisture Curve
HarvestMaster recommends testing the new moisture curve against other grain samples to confirm the
moisture.
1. Test the grain sample in a benchtop sensor by cycling sub-samples from the larger sample 3–5 times
and averaging the moisture readings. The averages are the Known Moisture.
2. Within the hour, take the sample out to the GrainGage. If you are calibrating outside, make sure the
sample stays in the shade.
3. Open Mirus.
4. Go to
Setup
>
H2 GrainGage
>
Moisture Curves
.
5. Select the moisture curve you want to test.
6. Tap the edit icon .
Gear icon