87R-4P Compressor System User Guide
6190900 (Rev B)
Site Requirements
Specifications
100 V
50-60 Hz
115 V
50-60 Hz
230 V
50-60 Hz
Electrical
Voltage
min./max.
90/110
103/127
207/253
Full load amps
(50 Hz)
5.0
4.8
2.6
Full load amps
(60 Hz)
5.8
5.3
2.9
Placement
• The system is intended only for indoor use in a climate-
controlled and dust-free environment.
• Do NOT install/operate the system in an enclosed area
where the ambient temperature could fall below 10 °C
(50 °F) or rise above 40 °C (104 °F).
• Maintain a minimum clearance of 30 cm (12 in) on the
top and sides of the system for service access and
cooling.
• Ensure that the system is placed on a surface with an
incline that does not exceed 10°.
Indicates the ON and OFF position for the
equipment power switch (system breaker)
When
ON
, the indicator light will illuminate
and current
WILL
be
supplied to system.
When
OFF
, the indicator light will
NOT illuminate and current
WILL NOT
be
supplied to the system.
I
o
= ON
= OFF
I
o
Pneumatic Connections
WARNING
The system should only be installed by qualified personnel
in a clean, dry, well-ventilated area on a firm surface with an
incline of less than 10°.
Equipment Alert: Verify that all leaks are sealed.
Air leaks are the primary cause of premature
system failures.
Connect the plumbing from the device air-inlet to the 1/4 in
BSPP air-output fitting on the back panel of the system.