29
8 The redox combination electrode
JUMO ecoLine Rd, 2 m fixed connecting cable and BNC connector
Type: 201010/54-22-65-2000-76-120/000
Sales No. 20/00424950
8.1
General handling
- Store in a dry room at a temperature between 10°C and 30°C.
Below 0°C the combination electrode can be destroyed by freezing of the
electrolyte!
- The sensor portion of the combination electrode should be protected
against drying out by using a protective cap filled with 3-mol KCl solution or
a storage holder.
Lengthy storage in de-ionized (distilled) water will result in a depletion of the
KCl within the combination electrode.
- It is neither necessary nor possible to top up the electrolyte for this
combination electrode (low-maintenance).
- Rinse the electrode carefully with de-ionized (distilled) water before making
a measurement.
- When measuring in different media, rinse thoroughly each time.
- Clean the combination electrode after use.
For contamination with proteins, a hydrochloric acid solution
(Sales No. 20/00307114) is recommended.
- The platinum tip of the sensor can be cleaned with ordinary scouring
powder (rinse well afterwards).
- If there is a malfunction of the combination electrode:
- check the function with a second reference combination electrode.
- replace the combination electrode.
- Solutions for calibration, cleaning and activation can be obtained from
JUMO or from laboratory suppliers.
Recommended
redox
combination
electrodes
Reference system: Ag/AgCl, electrolyte: KCl, 3 mol/liter
JUMO ecoLine Rd, 2 m fixed connecting cable and BNC connector
Type: 201010/65-22-65-2000-76-120/837
Sales No. 20/00424950
(use only this type for conversion to the hydrogen system “mV
H
” and rH
measurements!)
Operating life
The operating life of electrodes depends on the conditions in which they are
used. The operating life is at least 6 months under normal conditions and with
appropriate care and maintenance.
The hand-held meter is designed to be operated with electrodes that have a
silver/silver chloride (Ag/AgCl) reference system!
If a different type is used, the mV
H
measurement function will produce false
measurements!