3 Configuration
3.2
Select 2- or 3-point calibration
(only for pH measurement)
✱
Start configuration.
✱
Select the parameter “CAL”.
✱
Use the
or
key to select the type of calibration.
3.3
Select temperature unit (°C/°F)
✱
Start configuration.
✱
Select the parameter “Unit”.
✱
Use the
or
key to select the temperature unit.
3.4
Set temperature offset
✱
Start configuration.
✱
Select the parameter “OFFS”.
2-Pt: 2-point calibration.
The calibration of the pH combination electrode is carried out for two
points (one point in the neutral range, and one point in the acidic or
alkaline range).
3-Pt: 3-point calibration.
The calibration of the pH combination electrode is carried out for three
points (one point in the neutral range, and one point each in the acidic
and alkaline ranges).