4 Calibrating a pH combination electrode
18
bottom left, the measurement is stable.
✱
Press the
key to accept the measurement.
The 3-point calibration has now been successfully concluded, and the state
of the combination electrode will be displayed, see Chapter 4.5 “Display of
the electrode state”, page 20.
4.4
Without temperature sensor
Calibration
point 1
✱
Immerse the combination electrode in the first buffer solution, and stir.
✱
When the display has stopped blinking and the “OK” arrow appears at
bottom left, the measurement is stable.
✱
Press the
key to accept the measurement.
The hand-held meter will request that you enter a temperature.
✱
Enter the present temperature of buffer 1 with the
or
key.
✱
Press the
key to accept the measurement.
Rinsing
✱
Rinse the combination electrode with de-ionized (distilled) water and dry
off.
Calibration
point 2
✱
Immerse the combination electrode in the second buffer solution, and stir.
✱
When the display has stopped blinking and the “OK” arrow appears at
bottom left, the measurement is stable.
✱
Press the
key to accept the measurement.
The hand-held meter will request that you enter a temperature.
✱
Enter the present temperature of buffer 2 with the
or
key.
✱
Press the
key to accept the measurement.
Store