21
4 Calibrating a pH combination electrode
Show slope at
25°C
✱
Briefly press the
key again.
Permissible
range
-62 to -45 mV/pH
Optimum
-59.2 mV/pH
2
CAL
3
pH
mV
mV
H
rH
ATC
ok
CAL
AL Logg
pH
mbar
Slope -52.9 mV/pH
After the 2-point calibration, the slope (SL.) for the entire range will be shown.
After the 3-point calibration, the slope shown first is that for the acidic range
(SL. 1). If you press the CAL key again, the slope shown will be that for the
alkaline range (SL. 2).