6 Error and system messages

Error or 
system 
message

Cause

Remedy

General:

Low battery voltage. 
Functionality can only be ensured 
for a short time.

Fit a new battery.

Low battery voltage.
Operating from electrical supply: 
wrong voltage.

Fit a new battery.
Check power supply and replace if 
necessary.
If the error persists, then the hand-held meter 
is faulty.

No display
or
crazy 
characters

Low battery voltage.
Operating from electrical supply: 
power supply is faulty or wrong 
voltage/polarity.
Instrument error or fault.

Fit a new battery.
Check power supply and replace if 
necessary.
Remove battery or power supply – wait for 5 
seconds – reconnect.
If the error persists, then the hand-held meter 
is faulty.

Gone beyond the measurement 
range. Defect in combination 
electrode, temperature sensor or 
connecting cable.

Check combination electrode, temperature 
sensor and connecting cable, and replace if 
necessary.

Gone below measurement range.
Defect in combination electrode, 
temperature sensor or cable.

Check combination electrode, temperature 
sensor and connecting cable, and replace if 
necessary.

Instrument error.

Switch off the hand-held meter, and switch
on again after 5 seconds.
If the error persists, then the hand-held meter 
is faulty.

Only for temperature measurement (thEr)

No temperature sensor 
connected.
Temperature sensor is faulty.

Connect/replace the temperature sensor.

Value could not be
calculated.

Outside the permissible temperature
compensation range (0 to 105°C) or outside 
the permissible measurement range (Err.1 or 
Err.2). 

Only for pH measurement:

Содержание 202710

Страница 1: ...Hand held Meter for analytical measurement variables pH redox temperature Type 202710 and Type 202720 B 20 2710 0 Operating Instructions 2010 11 08 00461261 ...

Страница 2: ......

Страница 3: ...4 Set temperature offset 11 3 5 Set automatic power off 12 3 6 Adr 12 3 7 Manual temperature setting for operation without a temperature sensor 13 3 8 Manual pH setting for the rH measurement function 13 4 Calibrating a pH combination electrode 15 4 1 Preparation 15 4 2 Start 16 4 3 With a temperature sensor 17 4 4 Without temperature sensor 18 4 5 Display of the electrode state 20 4 6 Parameters ...

Страница 4: ...ination electrode 26 7 1 pH combination electrodes for various application areas 26 7 2 Cleaning 27 7 3 Operating life 27 7 4 General handling of pH combination electrodes 27 8 The redox combination electrode 29 8 1 General handling 29 9 Technical data 30 ...

Страница 5: ...d or not followed correctly Note This symbol is used when your special attention is drawn to a remark abc1 Footnote Footnotes are remarks that refer to specific points in the text Footnotes consist of two parts A marker in the text and the footnote text The markers in the text are arranged as continuous superscript numbers Action instruction This symbol indicates that an action to be performed is ...

Страница 6: ...t in safety or emergency stop installations or in applications where a malfunction of the device could lead to injury to personnel or significant material damage Please do not attach any equipment to socket 20 see Chapter 2 5 Connections page 9 Correct functioning and operational reliability of the instrument will only be assured when the permissible climatic conditions are observed see Chapter 9 ...

Страница 7: ...ecalibration 5 ATC arrow in the operating modes pH mVH or rH this indicates that the temperature sensor is connected and the automatic temperature compensation is active 6 OK arrow this indicates that the measurement was stable for a sufficiently long time 7 CAL arrow in pH operation this indicates that the instrument is in calibration mode 8 Subsidiary display Temperature measurement or operator ...

Страница 8: ...ode condition and calibration data press for 2 seconds starts the pH calibration 13 Set Menu press briefly Set for pH rH and mVH manual temperature entry if no temperature probe is connected Additionally for rH manual entry of the pH value press for 2 seconds set the operating mode Menu call up the configuration 14 Store Quit During measurement hold and store the present measurement The display sh...

Страница 9: ... socket has no function Please do not connect any equipment here 21 BNC socket Connection for pH or redox combination electrode 22 Mini DIN socket Connection for a Pt100 temperature sensor 4 wire connection A 2 wire connection is possible but there will be an additional measurement error depending on the lead length 20 21 22 ...

Страница 10: ...h a pH combination electrode mV Redox measurement with a redox combination electrode mVH As for mV The indicated value is converted to the standard hydrogen system temperature independent DIN 38 404 This requires the connection of the temperature sensor ATC or a manual temperature entry rH The rH value is calculated from the measured pH redox and temperature values The pH and temperature values ca...

Страница 11: ...oint calibration The calibration of the pH combination electrode is carried out for two points one point in the neutral range and one point in the acidic or alkaline range 3 Pt 3 point calibration The calibration of the pH combination electrode is carried out for three points one point in the neutral range and one point each in the acidic and alkaline ranges max 2 min 5 pH mV mVH rH ATC ok CAL AL ...

Страница 12: ...his amount to compensate for probe or measurement deviations Measured temperature minus offset displayed temperature Off The zero point shift is switched off 0 0 max 2 min 5 C pH mV mVH rH ATC ok CAL AL Logg pH max min 2 5 1 120 Automatic power off delay in minutes If no key is pressed on the hand held meter it will switch off after this time has elapsed Off Continuous operation appropriate when f...

Страница 13: ...he or key to set the temperature for the measured medium Confirm the entry with the key 3 8 Manual pH setting for the rH measurement function For exact measurement of rH the hand held meter must know the temperature and pH of the medium being measured The hand held meter can also measure the rH value for a liquid when the temperature and pH values are entered manually Start configuration Select th...

Страница 14: ...et the pH value Confirm the entry with the key The hand held meter now switches back to measurement mode 0 00 14 00 Setting range for manual entry of the pH value max 2 min 5 max min 2 5 pH mV mVH rH ATC ok CAL AL Logg pH mbar Store Quit 6 ...

Страница 15: ... electrode the protective cap contains 3 mol KCl Rinse the combination electrode and the temperature sensor if present with de ionized distilled water and dry off Buffer solutions have a limited usable life that is for instance seriously reduced by inadequate rinsing and drying of the combination electrode when changing solutions This can lead to calibration errors Use the freshest possible soluti...

Страница 16: ...ong press on the key is used to switch between various buffer series Buffer solutions can be obtained from laboratory suppliers 2 CAL 3 Values at 25 C 4 00 pH 7 00 pH pH mV mVH rH ATC ok CAL AL Logg pH mbar neutral calibration solution 7 00 pH Values at 25 C 01 68 pH A 04 01 pH C 06 87 pH D 09 18 pH F 12 45 pH G pH mV mVH rH ATC ok CAL AL Logg pH mbar neutral calibration solution 6 87 pH Input ran...

Страница 17: ...will be displayed see Chapter 4 5 Display of the electrode state page 20 Rinsing Rinse the combination electrode and the temperature sensor with de ionized distilled water and dry off Calibration point 3 only for 3 point calibration Immerse the combination electrode and the temperature sensor in the third buffer solution Buffers from the standard or DIN series will automatically be detected Enter ...

Страница 18: ...ill request that you enter a temperature Enter the present temperature of buffer 1 with the or key Press the key to accept the measurement Rinsing Rinse the combination electrode with de ionized distilled water and dry off Calibration point 2 Immerse the combination electrode in the second buffer solution and stir When the display has stopped blinking and the OK arrow appears at bottom left the me...

Страница 19: ... will automatically be detected When the display has stopped blinking and the OK arrow appears at bottom left the measurement is stable Press the key to accept the measurement The hand held meter will request that you enter a temperature Enter the present temperature of buffer 3 with the or key Press the key to accept the measurement Rinsing Rinse the combination electrode with de ionized distille...

Страница 20: ...od condition 90 30 The state of the combination electrode is satisfactory less than 30 The combination electrode is severely aged or contaminated If there is no improvement after cleaning see Chapter 7 2 Cleaning page 27 and a subsequent calibration then the combination electrode must be replaced 2 CAL 3 pH mV mVH rH ATC ok CAL AL Logg pH mbar The evaluation takes into account both the asymmetry a...

Страница 21: ...um 59 2 mV pH 2 CAL 3 pH mV mVH rH ATC ok CAL AL Logg pH mbar Slope 52 9 mV pH After the 2 point calibration the slope SL for the entire range will be shown After the 3 point calibration the slope shown first is that for the acidic range SL 1 If you press the CAL key again the slope shown will be that for the alkaline range SL 2 ...

Страница 22: ... held meter Configure the hand held meter for rH measurement see Chapter 3 1 Select measurement function pH redox mV redox mVH rH thEr page 10 The main display will show the rH value that has been determined The subsidiary display will alternately show the previously measured pH and the temperature For this measurement special care should be taken that both the pH and redox combination electrodes ...

Страница 23: ...nter the temperature Briefly press the key Use the or key to enter the pH value Briefly press the key The main display will show the rH value that has been determined The subsidiary display will alternately show the previously measured pH and the temperature The measurements for temperature and pH can also be manually entered in the hand held meter Set Menu 4 max 2 min 5 Set Menu 4 max 2 min 5 Sto...

Страница 24: ...hand held meter is faulty Gone beyond the measurement range Defect in combination electrode temperature sensor or connecting cable Check combination electrode temperature sensor and connecting cable and replace if necessary Gone below measurement range Defect in combination electrode temperature sensor or cable Check combination electrode temperature sensor and connecting cable and replace if nece...

Страница 25: ...n the combination electrode Repeat calibration Replace the combination electrode Always use the neutral buffer as the first solution Use fresh buffer solution The slope is too flat combination electrode is faulty buffer solution is exhausted Replace the combination electrode Use fresh buffer solution The slope is too steep combination electrode is faulty buffer solution is exhausted Replace the co...

Страница 26: ...bjected to chemical and mechanical wear and must therefore be replaced when the required parameters can no longer be upheld even after meticulous cleaning When using them it must be taken into account that some substances in aqueous solutions e g hydrofluoric acid HF can attack glass Chemicals may also react with the KCl electrolyte in the combination electrode and or lead to the diaphragm becomin...

Страница 27: ...re in a dry room at a temperature between 10 C and 30 C Below 0 C the combination electrode can be destroyed by freezing of the electrolyte The sensor portion of the combination electrode should be protected against drying out by using a protective cap filled with 3 mol KCl solution or a storage holder Lengthy storage in de ionized distilled water will result in a depletion of the KCl within the c...

Страница 28: ...he measuring function of the electrode is impaired for instance the display is very sluggish check the combination electrode for air bubbles check the function with a second reference combination electrode replace the combination electrode Solutions for calibration cleaning and activation can be obtained from JUMO or from laboratory suppliers ...

Страница 29: ...a pepsin hydrochloric acid solution Sales No 20 00307114 is recommended The platinum tip of the sensor can be cleaned with ordinary scouring powder rinse well afterwards If there is a malfunction of the combination electrode check the function with a second reference combination electrode replace the combination electrode Solutions for calibration cleaning and activation can be obtained from JUMO ...

Страница 30: ... for 4 wire Pt100 can also be connected as 2 wire Input resistance pH redox 1012 Ohm Nominal temperature 25 C Working temperature 0 to 50 C Relative humidity 0 95 r h no condensation Storage temperature 20 to 70 C Memory Min value Max value Hold memory Electrical supply 9 V battery type IEC 6F22 included in delivery Power input socket inner pin 1 9 mm dia pole for external supply 10 5 12 V DC use ...

Страница 31: ... temperature dependency is automatically compensated Manual buffers can be used Permissible electrode data asymmetry 60 mV slope 62 to 45 mV pH Sensor weighting 10 to 100 according to the calibration results Recommended redox combination electrodes Reference system Ag AgCl electrolyte KCl 3 mol liter JUMO ecoLine Rd 2 m fixed connecting cable and BNC connector Type 201010 54 22 65 2000 76 120 000 ...

Страница 32: ......

Отзывы: