8 Calibration
- Calibration must be enabled,
see Chapter 6.9.1 “Administrator levels”, page 39.
- The transmitter is in the measurement mode.
✱
Immerse the combination electrode in a buffer solution with a known ORP.
✱
Start the calibration (by pressing the
key, or via the Administrator level).
✱
Using the
key, start 1-point calibration.
Wait until the displayed value is stable; then continue with
.
✱
Set the displayed value to the value of the buffer solution using the
or
key; then continue with
.
The ORP of the sample solution is not dependent on temperature!