8 Calibration
✱
Start the calibration (by pressing the
key, or via the Administrator level).
✱
Using the
key, start 2-point calibration.
✱
With manual temperature input, set the temperature of the buffer solution
using the
or
key and confirm the selection with
.
✱
Wait until the displayed value is stable; then continue with
.
✱
Set the displayed value to the value of the first buffer solution (e.g. 4.00)
using the
or
key; then continue with
.
✱
Rinse, then dry the pH combination electrode.
✱
Immerse the pH combination electrode in the second buffer solution (e.g.
8.00).
CAL
PGM
PGM
PGM
PGM