8 Calibration
- Calibration must be enabled,
see Chapter 6.9.1 “Administrator levels”, page 39.
- The transmitter is in the measurement mode.
✱
Immerse the combination electrode in the first buffer solution with a known
pH value.
✱
Start the calibration (by pressing the
key, or via the Administrator level).
✱
Using the
key, start 3-point calibration.
With manual temperature input, set the temperature of the calibration solution
using the
or
key and confirm the selection with
.
The temperature of the three buffer solutions must be the same
and must remain constant during calibration.
CAL
PGM
PGM