Remote control
46
in commands:
Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
version
None
Number of software version (V X.xx)
status
none
Status message, error message (see page 47 )
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04
none
Setpoint temperature („SafeTemp“) of the excess
temperature protection
in_sp_00
none
Working temperature „t 1“
in_sp_03
none
High temperature warning limit
„t High“
in_sp_04
none
Low temperature warning limit
„t Low“
in_sp_07
none
Pump pressure stage
in_par_00
none
Temperature difference between working sensor and safety
sensor
in_par_01
none
Te - Time constant of the external bath.
in_par_02
none
Si - Internal slope
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_mode_05 none
Circulator in Stop/Start condition:
0 = Stop
1 = Start