Operator responsibility – Safety instructions
12
Eco- Series
SC 10000w
Working temperature ranges:
Standard unit
°C
+5 ... +35
Option Low Temp
°C
-20 ... +35
Option Low /HighTemp I
°C
-20 ... +80
Option Low /HighTemp II
°C
-20 ... +130
Temperature stability
°C
0.1
Absolute Temperature Calibration
±3
Heater wattage, Option H5 / H12
kW
5.0 / 12.0
Cooling capacity
°C
+20 0 -10
Medium ethanol
kW
10 5.0 2.5
Refrigerant
R404A, R452A*
Pump capacity P3 (Standard) / P4 ** (Option)
see table 1 page 13
Flow rate
P3 / P4
l/min at 0 bar
33 / 43
Pressure max.
P3 / P4
bar at 0 liters
3.5 / 4.4
Overall dimensions (WxDxH)
cm
59x67/112
Filling volume
liters
43 ... 60
Weight
kg 159
Ambient temperature
°C
5 ... 40
Mains power connection
V/ Hz
365-440V/3PNPE/50Hz
400/50/3 phases
Current consumption without heater/phase / P3 / P4
A
11 (400 V)
Current consumption 5 kW heater/phase / P3 / P4
A
17 (400 V)
Current consumption 12 kW heater/phase / P3 / P4
A
31 (400 V)
Mains power connection197-254 V/3PPE/60Hz
V/ Hz
208-230/60/3 phases
Current consumption without heater/phase / P3
A
15 (208 V) / 16 (230 V)
Current consumption without heater/phase / P4
A
17 (208 V) / 17 (230 V)
Current consumption 5 kW heater/phase / P3
A
25 (208 V) / 26 (230 V)
Current consumption 5 kW heater/phase / P4
A
26 (208 V) / 27 (230 V)
Current consumption 12 kW heater/phase / P4
A
44 (208) / 46 (230 V)
* at 400 V / 50 Hz
All measurements have been carried out at: rated voltage and frequency , ambient temperature: 20 °C
Technical changes without prior notification reserved.
** Pump P4- reduces cooling capacity by 0.3 kW