Appendix
95
Error code Description
Solution
•
Switch off the unit at the mains
switch, wait 4 seconds and then
switch the unit on again.
•
For water-cooled units: Check cooling
water temperature and supply.
•
If the fault has not been remedied,
contact Technical Service.
-1431
The minimum permissible current
consumption at the compressor has
been fallen short of.
•
Check mains voltage for nominal
voltage. The specified voltage
tolerance of the unit must not be
exceeded.
•
Check the mains cable of the
refrigeration unit for damage and
replace if necessary.
•
Check ambient temperature and
reduce if necessary.
•
Check CAN bus cable for damage and
replace if necessary.
•
If the fault has not been remedied,
contact Technical Service.