Initial operation
53
7.3 Switch on the device and set the language
The language must be set first, to prepare for important basic settings. When
switching on for the first time, alarm messages may be shown on the display.
The device is connected and ready for operation.
1. Switch the device on at the mains switch.
The software boots and starts the device. The display shows the device
name, voltage version and software version.
Since no temperature liquid has been filled yet, the low level alarm is
displayed.
Depending on the default setting, the high temperature cut-off alarm is
displayed.
2. Call up the
<Main menu>
.
3. Call up the submenu
<Settings>
then menu option
<Language>
.
The submenu
<Language>
lists all the languages installed on the device.
4. Select the desired language.
5. Turn the device off so that the alarm message is reset.
The language is set.
7.4 Set high temperature safety function
Before each new temperature application, the temperature must be set for the
tank and internal reservoir high temperature cut-off. Set a value that is at least
25 K below the flash point of the bath fluid being used. The surface temperature
of the bath fluid must not exceed the flash point at any time. An alarm is
triggered when the set value is exceeded.
Slotted screwdriver, size 3
The device is connected.
The language is set.
1. Switch the device on.
If no temperature liquid has been filled yet, the low liquid level alarm is
displayed.
Depending on the default setting, the high temperature cut-off alarm is
displayed.
2. Press the
[Display]
softkey to select the home screen that displays the high
temperature cut-off value.