Appendix
90
out_mode_04
x
Setting of the temperature control:
0 = Internal control
1 = External control
out_mode_05
x
Start/stop command of the device in remote control
mode:
0 = Stop tempering
1 = Start tempering
out_mode_08
x
Setting of the internal control dynamic:
0 = Aperiodic
1 = Standard
out_mode_11
x
Actuating variable source:
0 = Thermostat
1 = Serial
2 = Analog (EPROG)
System parameter Parameter
Setting
out_par_04
xxx
Setting the Cospeed control parameter for internal
control
out_par_06
xxx
Setting of the control parameter XP for internal
control
out_par_07
xxx
Setting of the control parameter TN for internal
control
out_par_08
xxx
Setting of the control parameter TV for internal
control
out_par_09
xxx
Setting of the control parameter XP of the
subordinate controller (cascade temperature control)
out_par_10
xxx
Setting of the control parameter XPU of the
subordinate controller (cascade temperature control)
out_par_11
xxx
Setting of the control parameter TN of the
subordinate controller (cascade temperature control)
out_par_12
xxx
Setting of the control parameter TV of the
subordinate controller (cascade temperature control)
out_par_13
xxx
Setting of the maximum permissible internal
temperature InternMax in cascade temperature
control
out_par_14
xxx
Setting of the minimum required temperature
InternMin in cascade temperature control