Menu functions
40
Should such a safety standard not be required, the AUTOSTART function
(automatic start mode) may be activated, thus allowing the start of the
instrument directly by pressing the mains power switch or using a timer.
Possible parameters:
on
- AUTOSTART on
off
- AUTOSTART off
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid any
hazardous situation to persons or property.
The instrument does no longer conform to N.A.M.U.R. recommendations.
Take care you fully observe the safety and warning functions of the instrument.
If the setpoint is given via the serial interface, no AUTOSTART is possible.
STAND-BY
>Stand-by input
External stand-by for emergency switch-off
(connector - see page 61)
Possible parameters:
no
- stand-by input is ignored
yes
- stand-by input is active
> Language
There are two options for the language of the DIALOG-DISPLAY (LCD):
German or English.
Possible parameters:
German
(deutsch)
English
(engl).
-OFF-
S xx.xx
ID 0
←
>IdentNo
There is no ident number for this unit.