Temperature System Magnum 91
67
11.4. Status messages
The instrument sends data (including error messages) only when the computer sends a query.
Status messages
Description
00 MANUAL STOP
Magnum in „OFF“ state.
01 MANUAL START
or
01 MANUAL START,DEGASING
Magnum in keypad control mode
or
Magnum in keypad control mode and degasifying active
02 REMOTE STOP
Magnum in „r OFF“ state.
03 REMOTE START
or
03 REMOTE START,DEGASING
Magnum in remote control mode.
Or
Magnum in remote control mode and degasifying active
11.5. Error messages
Error messages
Description
-01 NIVEAU LEVEL ALARM
Unterniveau-Alarm im internen Reservoir
-02 REFRIGERATOR ALARM
Defect temperature sensor in refrigeration circuit.
-03 EXCESS TEMPERATURE
WARNING
High temperature warning "
".
-04 LOW TEMPERATURE
WARNING
Low temperature warning "
".
-05 WORKING SENSOR ALARM
Working temperature sensor short-circuited or interrupted.
-06 SENSOR DIFFERENCE ALARM
Sensor difference alarm.
Working temperature and safety sensors report a temperature
difference of more than 50 K.
-07 I
2
C-BUS ERROR
Internal error when reading or writing the I
2
C bus.
-08 INVALID COMMAND
Invalid command.
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
Invalid command in current operating mode.
-10 VALUE TOO SMALL
Entered value too small.
-11 VALUE TOO LARGE
Entered value too large.
-12 TEMPERATURE
MEASUREMENT ALARM
Error in A/D converter.
Содержание 9410191.7
Страница 37: ...Temperature System Magnum 91 37...
Страница 74: ...Repairing the unit transport 74 Caution Transport...