Temperature System Magnum 91
73
12.1. Maintaining the cooling performance
To maintain the full cooling performance, clean the filter in the cooling water
entry from time to time.
•
Switch off the unit, disconnect mains power cable.
•
Interrupt the cooling water supply.
•
Disconnect the tubing from the inlet "IN" (14) and remove the dirty filter.
•
Clean the filter.
•
Replace the filter and reconnect the tubing.
•
Open the cooling water supply.
•
Check the tubing connection for tightness.
•
Unit is ready to operate.
13. Repairing the unit, transport
The unit is designed for continuous operation under normal conditions. Periodic
maintenance is not required.
Repairs
Before asking for a service technician or returning a JULABO instrument
for repair, please contact an authorized JULABO service station.
When returning the unit:
•
Empty completely and close tightly the connections with nuts and closing
caps.
•
Attach a short fault description.
•
Careful and appropriate packing is important.
•
During transport the unit has to stand upright. Mark the packing
correspondingly.
•
When returning a unit, take care of careful and adequate packing.
JULABO is not responsible for damages that might occur from
insufficient packing.
JULABO behält sich das Recht vor, im Sinne einer Produktverbesserung
notwendig gewordene Veränderungen technischer Art, die zu einer
einwandfreien Funktion beitragen, während des Reparaturvorgangs
durchzuführen.