Preparations
22
5.3.1. Tubing
Recommended tubing:
Metal tubing, triple insulated, M124x1,5 , Temperature range -100 °C ... +350 °C
Order No.
Length
max. pressure
at 20 °C
at 350 °C
8 930 261
1.0 m
6.0 bar
4.0 bar
8 930 262
1.5 m
6.0 bar
4.0 bar
8 930 263
2.0 m
6.0 bar
4.0 bar
8 930 264
3.0 m
6.0 bar
4.0 bar
Warning:
Tubing:
At high working temperatures the tubing used for temperature application and cooling water
supply represents a danger source.
A damaged tubing line may cause hot bath liquid to be pumped out within a short
time.
This may result in:
•
Fire
hazard
•
Explosion
hazard
•
Burning of skin
•
Difficulties in breathing due to hot atmosphere
Safety recommendations
•
Employ suitable connecting tubing.
•
Make sure that the tubing is securely attached.
•
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
•
Regularly check the tubing for material defects (e.g. for cracks).
Содержание 9410191.7