8.6. Limits
Int.Programmer
Inputs/Outputs
Limits
Interface
Temp.Sensor
Limits
>IntMax 400.00°C
IntMin -99.99°C
Band 30K
HeatingMax 100%
CoolingMax 100%
The limits IntMax and IntMin are only valid under external control (see 7.5.
Internal / external control). They restrict the temperature of the
internal bath to the desired maximum/minimum, also if the controller would
require a higher/lower temperature for the external system. As a result the
external setpoint may thus not be reached.
When operating the temperature system under external control, band
limiting is active. The preset value determines the maximum temperature
difference between the internal bath and the external load. This
adjustment possibility prevents sensitive equipment and temperature
devices from damage.
Heating and cooling power of the instrument are adjustable.
100 % corresponds to the values in the technical specifications of the
equipment.
•
Select the submenu "Limits" with enter
.
•
Select the desired option with the up/down cursor keys
.
A flashing digit indicates that a value needs to be entered.
IntMax
see temperature range LHxx
IntMin
see temperature range LHxx
Band
0 to 200 °C
HeatingMax
0 to 100 % in steps of 1 %
CoolingMax
0 to 100 % in steps of 1 %
•
Set the newly entered value by pressing enter
.
•
Return to the previous menu level by pressing escape
ESC
.
8.7. Interface
Interface
>Type RS232
Baudrate 4800
Parity no
Handshake softw
Inputs/Outputs
Limits
>Interface
Temp.Sensor
Interface
>Type RS232
Baudrate 4800
Parity no
Handshake softw
Interface
>Type RS485
Baudrate 4800
Parity no
Address 23
The interface parameters are set by selecting the submenu "Interface" on
the temperature system. Normally, this is a one-time-only adjustment.
•
Press
enter
to select the submenu "Interface".
•
Select the desired option with the up/down cursor keys
.
Enter
the desired value for the flashing digit.
Type
RS232 / RS485
Baudrate
4800/9600
Parity
none/even/odd
Handshake
software handshake/hardware handshake
Address
0 to 127
•
Press the P-key
to select the desired parameter and set with
enter
.
•
Return to the previous menu level with escape
ESC
.