![JTECH Medical COMMANDER ECHO Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/jtech-medical/commander-echo/commander-echo_manual_2028872064.webp)
© 2017 JTECH Medical MN022_M
62
®
EMC Guidance
WARNING:
Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this manual. Carefully
read the information contained in this section.
WARNING:
The use of portable and mobile RF equipment can affect the normal operation of
medical electrical equipment.
WARNING:
Making any modifications or using any accessories not specifically approved by JTECH
Medical may reduce immunity to electromagnetic interference or increase electromagnetic
emissions.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Commander Echo System is intended for use in the electromagnetic environment specified
below. The customer or user of the Commander Echo System should assure that it is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 2
The Commander Echo System must emit
electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be affected.
RF emissions
CISPR 11
Class B
The Commander Echo System is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A and D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
WARNING:
The Commander Echo instruments and Console should not be used while stacked on, or
adjacent to, other electrical or medical electrical equipment. If stacked or adjacent use is necessary,
all electrical equipment should be observed to verify normal operation.