![JTECH Medical COMMANDER ECHO Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/jtech-medical/commander-echo/commander-echo_manual_2028872046.webp)
© 2017 JTECH Medical MN022_M
44
®
Storage and Cleaning of Instruments
Notice:
The Commander Echo Console and its instruments are sensitive electronic devices, and should be
handled with care. If handled and cared for properly, the devices have all been designed and manufactured to
provide years of accurate and reliable use. Avoid dropping, banging, or other impacts to the instruments. Use
only within designated operating temperatures.
Storage
When Commander Echo instruments are not in use, store them in the protective carrying case.
1 .
If instruments are stored for periods longer than 30 days, recharge batteries before using.
2 .
Store in an area with a controlled temperature within the recommended temperature range. (see
temperature and humidity limits in “Technical Specifications” on page 51)
Important Notice:
Never store Commander Echo instruments in an automobile, except when
transporting them, regardless of the season.
Cleaning, Disinfection, and Sterilization
Warning:
Keep dry, do not immerse any part of any Commander Echo Console, Echo instrument,
Console transceiver, receiver, or accessory in water or any other fluid.
Commander Echo Console, instruments and accessories are non-sterile devices and are not compatible
with sterilization techniques such as autoclave.
•
Do not autoclave Echo instruments or accessories.
•
Do not clean with abrasive materials.
•
Do not clean with solvents, or disinfectant not approved by JTECH Medical.
Recommended Cleaning of Commander Echo Console and Instruments
•
Wipe surfaces with a soft, dry cloth.
Recommended Alternative Method for Cleaning Echo Instruments and Accessories.
•
CaviWipes disinfecting towelette
◌ Use only as directed, following all instructions and warnings on the CaviWipes packaging.
Warning:
Follow all directions as specified by the CaviWipe product being used. Allow devices to dry
before use.
Preventative Inspection
Echo devices have no associated methods of preventative inspection.
Maintenance
Echo devices are not user serviceable and do not include a service manual.
For any service including maintenance, repair, and calibration, contact JTECH Medical Customer Support at
1-800-985-8324 or 1-801-478-0680