![JTECH Medical COMMANDER ECHO Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/jtech-medical/commander-echo/commander-echo_manual_2028872062.webp)
© 2017 JTECH Medical MN022_M
60
Technical Specifications
Operating Temperature, Humidity, and
Atmospheric Pressure
10 %
90 %
50 kPa
106 kPa
Transportation and Storage Conditions
10 %
100 %
50 kPa
106 kPa
Internal Power Source
Non-User serviceable,
Lithium Polymer battery
Recharge time
Four(4) - six(6) continuous hours of charging following initial charge
Battery conservation
Instruments transition to “off power” state when not in use
Specified Power Supply,
(Battery Charger)
JTECH Medical PN: PW006 [Protek Power PMP15M-10-B6] or
JTECH Medical PN: PW008 [Globtek, Inc. WR9QA3000CCP-N-MED]
Type of Operation
Continuous
Type of Protection against electric shock
Internally powered equipment
Degree of protection against electric shock
Type B equipment
Protection against harmful ingress of water
IPX0 - ordinary equipment
Degree of Safety of Application in the presence
of flammable mixtures
This device is not suitable for use in the presence of flammable anesthetic mixture
with air, or with oxygen or nitrous oxide
This device contains:
FCC ID: OUR-XBEE
Model XBee-Radio, Canada IC: 4214A-XBEE
Japan ID: 005NYCA0378
RF operating distance
3 meters (9.8 feet) from console, indoor environment
RF frequency
2.4GHz wireless frequency
Symbols
IPX0
Medical Device Class
Class I measuring medical device per Annex IX of Council Directive 93/42/EEC
Class I medical device per CFR Title 21 Part 888 Supbart B
Class II medical device per Rule 10 of the Canadian Medical Devices Regulations
Dimensions
Weight
Maximum Force Input
Operating Temperature, Humidity, and
Atmospheric Pressure
10 %
90 %
50 kPa
106 kPa
Transportation and Storage Conditions
10 %
100 %
50 kPa
106 kPa
Type of Operation
Continuous
Input Voltage
100-240 VAC
Input Frequency
50-60 Hz
Input Current
0.5 A (rms) for 115 VAC
0.3 A (rms) for 240 VAC
Enclosure Leakage Current
100 uA max. @ 264 VAC, 63 Hz
Maximum Output Voltage/Current
5 V / 3 A
Maximum Output Power
15 W
Type of Protection against electric shock
Class II
Manufacture, Model
JTECH Medical PN: PW006 [Protek Power PMP15M-10-B6]
Protection against harmful ingress of water
IPX0 - ordinary equipment
Degree of Safety of Application in the presence
of flammable mixtures
This device is not suitable for use in the presence of flammable anesthetic mixture
with air, or with oxygen or nitrous oxide
Symbols
IPX0
Dimensions
7.8 cm x 5.2 cm x 3.5 cm (3.1 in x 2.0 in x 1.4 in)
Weight
0.21 kg (0.38 lb)
Specified Power Supply
(Battery Charger)
(PW006*)
Used with the Daisy Chain for charging your Echo devices.
*Power supply connection may vary according to the country shipped to
JTECH Marketing <marketing@