INSTRUCTION MANUAL CODE 80243 REV A
January/11
(It can be modified without notice)
Page: 51
J
.P. SELECTA s.a.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 93 770 08 77 Fax 34 93 770 23 62
e-mail: [email protected] - website: http://www.jpselecta.es
lamp. (Fig. 49)
When the used time of D2 lamp is higher than 1000 hours, replacing a new one is
strongly recommended.
When the used time of W lamp is higher than 1000 hours, replacing a new lamp is
strongly recommended.
When you have replaced a new lamp, you’ll have to go into the lamp life
setting screen, and move the cursor on the “Reset W Life” or “Reset D2 Life” by
pressing
[»]
and
[«]
followed by
[ENTER]
to reset them into zero.
7.8 Load Default
Move the cursor onto
Load Default
by pressing
[»]
and
[«]
in utility setting screen,
and then press
[ENTER].
(Fig. 50)
If you want to default the parameters, choose “Yes” in the lower screen of Fig. 50.
If not, you should choose “No”.
Press
[RETURN]
to exit without any choice.
7.9 Lamp Change
Move the cursor on “
Lamp Change”
by pressing
[»]
and
[«]
in utility setting screen,
then press
[ENTER]
. (Fig. 51)
Input the new wavelength you want by pressing the numeric keypad followed by
[ENTER]
.
Lamp change wavelength means the point at which W lamp and D2 lamp exchange.
The range of it is 300 - 400, other value is invalid.
7.10 Version of Instrument
Move the cursor onto
Version
by pressing
[»]
and
[«]
in the utility setting screen,
and then press
[ENTER]
. Then you will find the detailed messages of the instrument.
Figure 49
Figure 50
Figure 51