www.joerns.com | [email protected]
Joerns
®
Bariatric Beds
Bari10A Model
14
Scale System
Functions and
Weighing Procedures
Caution:
5th Wheel must be disengaged to obtain
an accurate resident/patient weight.
0.0 Zero
With all bed linens in place and without resident/
patient in bed, press
Zero
button. Displayed
message will read “PLEASE WAIT - HANDS OFF”.
Be sure nothing touches weight-sensitive portion
of integrated Scale System while performing
zeroing function. In a few seconds, display will
indicate “WEIGHT = 0.0 LB”.
Lb/Kg
Weight data may be viewed in either
pounds
or
kilograms
. Pressing this button allows you to
switch between display modes quickly and easily.
Both modes will display resident/patient’s weight
within 0.1 units of resolution (0.1 Lb. or 0.1 Kg.).
Freeze/Resume
Freeze/Resume
mode button provides
customized ability to adjust resident/patient’s
comfort by adding/ removing a pillow or blanket
or ancillary equipment, without affecting
resident/patient’s displayed weight.
Press
Freeze/Resume
button. The display will
read “FREEZE = (resident/patient’s weight).”
Make all necessary adjustments to bed or
resident/patient. When done, press
Freeze/Resume
button again to resume
normal weighing.
Weight Change Mode
This mode is used to monitor resident/patient’s
weight changes.
Note:
To enable this function
initially, press
Weight Change Mode
button with
resident/patient in bed and then press
Zero
button
to zero out weight. Display will read
“ WT CHG = (resident/patient weight)”.
Any variation in weight will now be viewed as a
“+” or “-” figure relative to
zero
.
Weight Mode
If display shuts off before you can view weight data,
simply press the
On
button to refresh display and
view resident/patient’s weight. Display will indicate
“WEIGHT = XXX.X LB”(or KG).
Caution
: After initialization process is complete and
Scale System is still in
Weight Change Mode
,
do
not
press
Zero
button
again
, or you will lose all
stored resident/patient’s weight history.
Pressing
Weight Mode
button will switch back to the
Weight Mode
allowing you to view resident/patient’s
total weight again.
Caution
: If you
do not
press
Freeze/Resume
button before adding/removing items from bed,
the weight of those items will be reflected in the
displayed weight, which will be hard to adjust back to
the correct stored weight.
Note:
You can correct any weight discrepancy
described above, by using
Weight Mode
(for
increases) and
Weight Change Mode
(for
decreases) buttons, to manually adjust displayed
weight reading. You will need to know the weight of
added/removed items to add/subtract from
total weight.
If
Freeze/Resume
mode is left unattended for
more than three minutes, display will read “LEFT IN
FREEZE - REFER TO MANUAL.” You must press
Freeze/Resume
button to exit this mode.
On
Button refreshes display or initiates power for
scale use.
0.0
zero
Freeze
Resume
Weight
Change
Mode
On
Note:
Material -
7 mil GLOSS POLYESTER - see finish notes below
Emboss Tool - 0.015”- 0.010” pillow emboss (for metal domes)
6 Buttons up to edge of and including the white region (ellipses) -
Note - Two Hidden buttons - (flat non tactile) see sheet 2of 2
Selective Hardcoat to be Velvet except in region of 6 keys (matching emboss)
and Display Window
Key embosses (six) to be Gloss
Adhesive Mask - selective adhesive tool
No adhesive behind clear window
Adhesive 3M part no 7959MP
Display window to be Gloss
See sheet 2 of 2 for dimensional layout, circuit layout, tail layout
and more specifications
Edge Bleed to be added
Selective adhesive die
on its own layer -
PMS 2715 C
Purple
PMS 425 C
Dark Gray
PMS 360 C
Green
PMS 429 C
Light Gray
OPAQUE
WHITE
Screened colors
Lb/Kg
1
.8
6
"
7.42"
Weight
Mode
.06 R. TYP. 4
NOTE - SEE SHEET 2 OF 2
FOR TWO HIDDEN KEYS
Rev
Part #:
Page 1 of 2
216-2268
Joerns Bari 10A Bed
Scale Membrane Keypad
INIT
Scott Technology, LLC
1 Jacks Hill Road
Oxford, CT 06478
Scott
Init ________/Date_________
Nexus
Init ________/Date________
Date
ECO
JP
00430
B
03/27/07
Scott
Init ________/Date_________
03/20/07
JP
-
A
Figure 21
Содержание Bari10A
Страница 12: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 12 F Figure 18...
Страница 20: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 20 Appendix A Actuator Wiring Diagram...
Страница 21: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model Appendix B Scale System Wiring Diagram 21...
Страница 22: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 22 Notes...