background image

www.joerns.com | [email protected]

Joerns

®

 Bariatric Beds

Bari10A Model

10

Mattress Support Platform    

Width Extenders 

The four Mattress Support Platform (MSP) 
sections each have a 6" Width Extender on 
both sides of the bed (Figure 12).

Release Width Extender by pulling latch 
handle along the bottom edge of each MSP 

section (Figure 13). 

Note:

 Width Extenders 

lock in either full open or full closed position 
and should 

only

 be used in these positions. 

Side Rails move with the Width Extenders.

Caution:

 Extend all four sections on each side 

prior

 to adjusting bed position or Side Rails.

Warning:

 Use Width Extenders only after the 

resident/patient is safely positioned in the bed. 
Failure to do so could result in injury.

Mattress Support Platform    

Length Extender 

Mattress Support Platform (MSP) Foot Section 
has an 8" Length Extender, as does the foot 
end of the bed frame. (Figure 14). 

Both MSP Length Extender and the Frame 
Length Extender latch in the normal and 
extended positions. 
Using the MSP Length Extender is a two-step 
process: 

Slide the Frame Extender out first, by 

pulling out the latch knob on the side of 

the bed frame (Figure 14). Begin carefully 

sliding the extender out and release the 
latch knob as soon as the extender starts 
to slide so that it will automatically latch 
again at the extended position. Slowly pull 
the extender out until the latch pin clicks 
into its detent. 
Reach under the center of the MSP Foot 
Section and locate the MSP Extender 

latch knob (Figure 15). Pull the knob out to 

unlatch the Foot Section Extender. Begin 
carefully sliding the extender and release 
the latch knob as soon as the extender 
starts to slide so that it will automatically 
latch again at the extended position. 
Slowly pull the extender out until the latch 
pin clicks into its detent.

Warning: Possible Equipment Damage. 

Do 

not slide Frame Length Extender past its detent. 
Damage to the scale cable will result.

1.

2.

1.

2.

3.

a.

b.

Figure 12

Figure 15 - Photo taken from angle  
 

underneath bed

Figure 14 - Arrow indicates Length  
 

Extension Latch

Warning: Possible Equipment Damage.

 Prior to 

use of the MSP Length Extender, the Frame Length 

Extender must be fully extended and latched. Failure 
to do so will result in scale connector damage.

Figure 13

Содержание Bari10A

Страница 1: ...Joerns Healthcare User Service Manual Joerns Model Bari10A Bariatric Bed leading bed evolution To avoid injury read user s manual before using bed...

Страница 2: ...not be placed in an oxygen enriched environment such as an oxygen tent Use of electrical circuits in an oxygen enriched environment could result in a fire hazard 2 Warning Possible Shock Hazard Unplu...

Страница 3: ...maneuverability performance push the bed from the head end To avoid injury to the resident patient and or user do not attempt to move the bed laterally with the steer mode engaged The steer wheel can...

Страница 4: ...f long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance...

Страница 5: ...ra 8 length for taller residents patients 18 3 4 to 33 3 4 Travel Range Integrated Scale System for resident patient weight monitoring Head and Foot End Side Rails Assist Devices Side Rails on Head an...

Страница 6: ...Lock for all functions to keep bed from accidentally operating using battery backup Note To avoid electric shock always plug bed into properly grounded hospital grade receptacle Actuator battery backu...

Страница 7: ...port Platform Release Lever Up Hold Emergency Release handle Up until Mattress Support Platform lowers to desired position Release handle if you need to stop descent 1 1 1 2 7 Figure 4 Bari Selective...

Страница 8: ...s rotate switch until it clicks into position Side Rails Assist Devices Bari10A has one Head end Side Rail and one Foot end Side Rail on both sides of bed Figure 7 Side Rails are attached and operated...

Страница 9: ...owest position with the siderails fully raised and locked when moving the bed with a resident patient in it For best maneuverability performance push the bed from the head end Warning 5th wheel must b...

Страница 10: ...p process Slide the Frame Extender out first by pulling out the latch knob on the side of the bed frame Figure 14 Begin carefully sliding the extender out and release the latch knob as soon as the ext...

Страница 11: ...t to lock all four casters individually to prevent unintended bed movement Always plug bed back into power outlet so that it is continuously charging Accessories Mattress Bari10A Bed may be equipped w...

Страница 12: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 12 F Figure 18...

Страница 13: ...ng Conditions Ambient Temperature 20 C to 30 C Relative Humidity 10 to 95 Non Condensing Storage and Shipping Conditions Ambient Temperature 30 C to 50 C Relative Humidity 10 to 95 Non Condensing 1 2...

Страница 14: ...will lose all stored resident patient s weight history Pressing Weight Mode button will switch back to the Weight Mode allowing you to view resident patient s total weight again Caution If you do not...

Страница 15: ...12 months or as your maintenance policy requires Electrostatic Discharge ESD Integrated circuits and semiconductors on printed circuit boards may be damaged by electrostatic discharge ESD Be sure to...

Страница 16: ...teries 1 2 3 1 2 3 4 Battery Replacement If Scale System batteries need replacement display will indicate CHANGE BATTERIES Remove Battery Cover from Foot Panel to access Scale System batteries Figure...

Страница 17: ...Use Super Lube synthetic grease or equivalent Oiling Points Keep pins and bolts oiled sample indicated by arrows in Figure 24 Oil hinge points sample indicated by arrows in Figure 25 Grease Points For...

Страница 18: ...d may be damaged Check cord for damage If cord is damaged discontinue bed usage until cord is replaced 1 2 3 4 If Adjustment Only partially occurs stops and beeps Battery may be low Plug bed into powe...

Страница 19: ...ctions may require two hands to apply equal pull force Contact Joerns Healthcare Technical Support at 800 966 6662 1 2 Mattress Support Platform sections do not articulate Pendant Control operation Se...

Страница 20: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 20 Appendix A Actuator Wiring Diagram...

Страница 21: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model Appendix B Scale System Wiring Diagram 21...

Страница 22: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 22 Notes...

Страница 23: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 23 Notes...

Страница 24: ...mal use These parts such as DC batteries are not covered under the one year warranty but are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to...

Отзывы: