d. Check structural parts for any cracking
(especially welds), bending or hole
enlargement.
e. Check wiring for proper connections and
damage (fraying, kinking or deterioration).
Report any damage to Joerns Healthcare
Technical Support at 800-966-6662.
f. Check Actuators for correct mounting at
attachment points and insure all related pins
are mounted securely and properly to frame.
Actuators are not serviceable.
If any discrepancies are noted during
inspection, they must be corrected
before continuing bed usage. Major
service and repair must only be
performed by authorized Joerns
Healthcare Technical Support, which
you may contact at 800-966-6662.
General Cleaning
For low level disinfecting, wipe surfaces with a
properly prepared detergent/disinfectant.
Do not use abrasive cleaners. Use damp rag
only when cleaning around electrical connectors
and components.
Actuator surfaces must be cleaned of dust and
dirt regularly. Do not attempt to clean exposed
piston rods. Never spray directly on Actuator
with high-pressure cleaners.
Storage
Store in a clean dry area.
Plug bed Power Cord into grounded hospital-
grade electrical outlet. Actuator Control System
backup batteries maintain the longest life if kept
charged and are not allowed to fully discharge.
Failure to keep Bari10A bed plugged in, will
discharge backup batteries and shorten their life.
If storing bed for periods longer than three
months, remove all Scale System batteries.
1.
2.
3.
1.
2.
3.
4.
Battery Replacement
If Scale System batteries need replacement,
display will indicate “CHANGE BATTERIES.”
Remove Battery Cover from Foot Panel to
access Scale System batteries (Figure 23).
Then, remove
all six
batteries (dispose used
alkaline batteries per local codes), and replace
with alkaline D cell batteries.
Printed circuit board vulnerable to electrostatic
discharge (ESD). Avoid touching circuit board
when changing batteries. Battery orientation is
illustrated on battery holder.
Cleaning, Inspection
and Maintenance
Warning
: Disconnect Bari10A Power Cord from
wall outlet before cleaning.
Warning
: Do not use power wash or high pressure
cleaner on
any
Bari10A part.
Warning
: Lock all four Casters when cleaning,
inspecting or performing any maintenance of the
Bari10A. Lower all Mattress Support Platform
sections unless necessary for positioning for
service access. Place all Selective Function
Lockout switches in
locked
position. Disconnect
Power and Pendant Cords.
Caution:
It is critically
important to lock
all four
casters individually.
To achieve maximum usage and longest life
for your Bari10A bed, follow all warnings and
cautions within this manual and maintain
your bed with care, as outlined below. The
maintenance required will be dictated by
your bed’s usage and care, which should be
inspected every six months.
To optimize bed longevity:
a. Keep bed clean.
b. Observe bed usage and operation and
report any discrepancies or problems to
Joerns Healthcare Technical Support at
800-966-6662.
c. Check for loose bolts, nuts or pins, and
report problems. After a week of
use, check all bolts for any looseness, and
make sure all pins are in their
normal location and fastened.
1.
2.
1.
2.
Содержание Bari10A
Страница 12: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 12 F Figure 18...
Страница 20: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 20 Appendix A Actuator Wiring Diagram...
Страница 21: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model Appendix B Scale System Wiring Diagram 21...
Страница 22: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 22 Notes...
Страница 23: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 23 Notes...