7
BOR-ON 2008-10-07
joined together with other materials in two or
more layers, without having to expose the
material on the opposite side.
2.3 Attach the lip clamp
Attachment
Be
fore attaching the lip clamp, make sure that
the surface to be tested has been completely
cleaned and that paint, coatings and dirt have
been ground off.
Thereafter, attach the lip clamp and fasten it by
tightening the threaded shaft with a 13 mm ring
spanner, until the full clamping force has been
attained.
The material is now ready to be tested with the
Material Analyser.
2.4 Material testing
Activate the Material Analyser
To activate the Analyser, move the front spring-
loaded clamping ring backwards until a click is
heard. When this has been done, the Analyser
is ready to carry out the material test.
Positioning the Analyser
Insert the analyser into the lip clamp. Make sure
that the analyser is in contact with the material
to be tested.
Testing
To carry out the test, move the front clamping
ring forwards, in order to release the test
mechanics of the analyser. A "click" is heard
from the analyser and the display shows the
measured material value.
Always measure with the Analyser vertical or
horizontal – Never
with an inclination greater
than 90°.
IF A NEW TEST IS TO BE CARRIED OUT ON
THE SAME MATERIAL,
NOTE!
ALWAYS CHANGE
THE POSITION OF THE RETAINER. NEVER
MAKE A SECOND TEST IN THE SAME
PLACE, SINCE THE MEASUREMENT VALUE