1.4.4 Medical regulations
The tester must not be operated by a person with a heart attack or with a pacemaker.
1.5 Test safety procedures
Never use the tester on a live circuit board or device! !!
The ground wire of the tester must be connected in accordance with regulations. When
connecting the test line, you must first connect the terminal under test on the tester to the object
to be tested. The high-voltage test lead can only be plugged into the high-voltage output before
testing. When holding the high-voltage test lead, it must be held in an insulated part, and must
not be held on a conductive body. The operator must be able to operate completely
independently. The switch and remote control switch cannot be controlled by others. The remote
control switch should be placed in a fixed position when not in use.
During the test, never touch the test object or anything connected to the test
object.
1.6 The following safety points must be kept in mind
·
Non-qualified operators and unrelated personnel should stay away from the high-voltage test
area.
·
Always maintain a safe and orderly condition in the high-voltage test area.
·
Do not touch the test object or any objects connected to the test object during the high voltage
test.
·
In case of any problems, turn off the high voltage output and input power immediately.
·
After the DC withstand voltage and insulation resistance test, the discharge operation must be
performed before the work of removing the test lead.