show:
Insulation resistance test
7. Voltage breakdown (BREAK)
If the leakage current of the object under test during the AC / DC withstand voltage test is far
beyond the range that the instrument can measure, and the arc current is far beyond the normal
value that the instrument can measure, it will be judged by the program as resistant. The test
failed due to a crash, and the LCD will display:
AC withstand voltage test
DC withstand voltage test
ACW BREAK
XXX.XS
or
DCW BREAK
XXX.XS
MX
X.XXKV
XX.XXmA
MX
X.XXKV
XX.XXmA
8. Overcurrent failure (OVER)
If the output current of the DUT during the AC / DC withstand voltage test exceeds the normal
output range of the instrument, it will be determined by the program as a test failure caused by
overcurrent, and the display will show:
AC withstand voltage test
DC withstand voltage test
ACW OVER
XXX.XS
or
DCW OVER
XXX.XS
MX
X.XXKV
>12 mA
MX
X.XXKV
>6 mA
Ground resistance overcurrent failed (OVER)
If the output current of the DUT during the ground resistance test exceeds the normal output
range of the instrument, it will be judged by the program as a test failure caused by overcurrent,
and the display will show:
GND OVER
XXX.XS
MX
>25A
XXXm
Ω
9. Passed the test (PASS)
If there is no abnormal phenomenon in the whole process of AC / DC withstand voltage or
insulation resistance test, the test object is deemed to pass the test, and the display will show:
AC withstand voltage test
DC withstand voltage test
ACW PASS
XXX.XS
or
DCW PASS
XXX.XS
MX
X.XXKV
XX.XXmA
MX
X.XXKV
XX.XXmA
Insulation resistance test
Ground resistance test
GND PASS
XXX.XS
MX
XX.XXA
XXXm
Ω