Please use the "+" or "-" key on the panel to input the standard value of the standard ammeter
into the calibration program, the unit is mA. Please confirm the number is correct before
pressing the "EXIT" key to close the output and save the input data. If the data is not saved,
press the "Reset" switch to turn off the output.
9.7 Correction of DC withstand voltage 2mA
Press the "SET" key, the program will enter the DC withstand voltage 2mA correction mode, and
the display will show:
Please connect a standard DC ammeter with a resistor of about 500K
Ω
/ 2W in series first, and
then connect it between the
“
H.V.
”
and the
“
terminal under test
”
of this instrument. Please
press the "Start" key. At this time, the calibration program of this instrument will automatically
output a current of about 1000VDC / 2.000mA, and the display will show:
Please use the "+" or "-" key on the panel to input the standard value of the standard ammeter
into the calibration program, the unit is mA. Please confirm the number is correct before
pressing the "EXIT" key to close the output and save the input data. If the data is not saved,
press the "Reset" switch to turn off the output.
9.8 Insulation resistance voltage correction:
Press the "SET" key, the program enters the insulation resistance voltage correction mode, and
the display will show:
Connect a standard voltmeter that can measure 1000VDC to
“
H.V.
”
and
“
terminal under test
”
on this instrument. If the standard voltmeter specifically indicates
“
high and low
”
, please
connect the high-end H. This instrument V. To the terminal, connect the low end to the "tested
end" of the instrument to avoid inaccurate or damaged voltmeter. Then press the "Start" key. At
this time, the calibration program of this instrument will automatically output a voltage of about
1000VDC, and the display will show:
CAL DCW I2 = 2.000mA
<TEST> to Calibrate
CAL DCW I2 = XX
.
XXmA
<EXIT> to Save
CAL IR V = 1000V
<TEST> to Calibrate
CAL IR V = XXXXV
<EXIT> to Save