22
HI-Limit = XXXXM
Ω
Range: 0-2000 0=OFF
LO-Limit = XXX.XM
Ω
Range: 1.0-999.9
Insulation resistance test
Ground resistance test
or
or
Please use the "+" or "-" key on the panel to enter the upper limit value to be set. If the upper
limit of the insulation resistance is set to "0", the program does not determine the upper limit of
the insulation resistance.
6. Leakage current or insulation resistance lower limit (LO-Limit) setting
After the leakage current or insulation resistance upper limit setting of AC / DC withstand
voltage test is completed and the
“
SET
”
key is pressed, the program will enter the leakage
current or insulation resistance lower limit setting mode or ground resistance lower limit setting
of AC / DC withstand voltage test. Mode, LCD display will show:
AC withstand voltage test
DC withstand voltage test
or
Insulation resistance test
Ground resistance test
or
or
Please use the "+" or "-" key on the panel to enter the lower limit value to be set.
7. Ramp Time setting
After the AC / DC withstand voltage current leakage limit is set and the
“
SET
”
key is pressed,
the program will enter the ramp-up time setting mode.
AC/DC withstand voltage test
Note: This function is not available in the insulation and ground resistance test. The program
will automatically skip this setting and go directly to the next setting. Please use the "+" or "-"
key on the panel to enter the slowly rising time to be set, the unit is second.
8.Dwell Time setting
After the ramp-up time setting of the AC / DC withstand voltage test is completed and the
“
SET
”
key is pressed, the program will enter the test time setting mode, and the LCD will
LO-Limit = X.XXmA
Range: 0.00-12.00
LO-Limit = X.XXmA
Range: 0.00-6.00
Ramp Time= XXX.XS
Range: 0.1-999.9
HI-Limit = XXXm
Ω
Range:1-300
LO- Limit = XXXmA
Range: 0-300